Release 24.12.11
New Features
Reference Number | Details |
---|---|
3612 | Added show trends button in finalized reports so that trends may be accessed directly from the report. |
3735 | Updated Somatom Quality Constancy PDF Import to support imports from Siemens Confidence Somatom model. |
3736 | The IC Profiler PDF Import will now display NaN (Not a Number) for penumbra values on FFF beams where there are no results. This allows the test to show as complete for older IC Profiler equipment that does not support FFF beams. |
3738 | For the Catphan analysis, reduced the overestimation and variability of slice thickness measurements for wire ramps. The three main differences are:
Testing showed these to produce more accurate slice thickness and less variation between the module's four wires. |
Reference Number | Details |
---|---|
3713 | Fixed a Machine Log issue where changing the specified machine for an incident to a machine that did not have the same custom fields resulted in the fields being shown. |
3729 | Resolved issue with Ion Chamber/Electrometer with disabled calibration factors displaying as available in schedule settings. |
3734 | Fixed issue for Catphan analysis where the sensitometry plug ROIs were not shown correctly on the report graphic. |
Release 24.11.27
New Features
Reference Number | Details |
---|---|
3715 | Added support for Siemens Biograph 64 Vision 600 PET CT to the Somatom Daily QA PDF Import. |
3710, 3639 | Added support for importing Siemens PET Syngo PDF files. To support the new analysis a new built in machine type, PET, was added. |
3717 | Added Help Request link to the main navigation pane so that users could link to the support page. |
3695 | Added schedule setting for IC Profiler analysis to account for whether or not the top plate is used. |
3698 | For the DSS (Decision Support System), which is still in beta testing, adjusted SPC limits to use the wider of control limits and action limits so that both the action and control limits are within view. |
Reference Number | Details |
---|---|
3714 | When editing a numeric table, a warning will be displayed if there are preset values for that table in the schedule settings. In addition if changes are made to a numeric table such that the schedule settings no longer match the table, we trap any errors to prevent JavaScript from crashing. |
3708 | Resolved issue with users were unable to preset lookup table meta-items for custom numeric tables in the schedule settings. |
3718 | Corrected sort order for file and image processing tests where the results where sorted by meta-items in the schedule settings. |
3722 | Fixed issue where empty custom numeric tables with no calculation columns were showing as complete in new QA. |
3724 | Resolved issue where settings in the Excel Spreadsheet and Upload Control Display sections were not copied when copying a schedule. |
Release 24.11.13
New Features
Reference Number | Details |
---|---|
3643 | Clinic hours are now supported at the account or site level when calculating schedule due dates, time to sign-off completion times in periodic reports, and downtime in machine logs. |
Reference Number | Details |
---|---|
3660 | Fixed issue with IC Profiler PDF import where tests displaying as complete in new QA despite one or more results are missing. The report rendered correctly. |
3704 | Corrected issue for custom deviation tests where the rollover point was always set to 360 when rendering in reports. |
3707 | Resolved issue in schedule settings where the frequency incorrectly displayed as ad hoc for yearly schedules. |
Release 24.10.30
New Features
Reference Number | Details |
---|---|
3658 | Improved the loading time of Report List screen. |
3689, 3696, 3691 | For the Magphan RT analysis, added contrast tests for the optional center section. The new tests are Contrast Sphere Signal Measurements, Background Fit, Background-Normalized Signa, Contrast Pairs, and Relative Contrast. There are new schedule settings to support these new tests. |
3693, 3697 | For Magphan analysis, added a test that makes the analysis log file available for download in a report. |
3694, 3692 | For the S162 analysis, added contrast tests for the optional center section. The new tests are Contrast Sphere Signal Measurements, Background Fit, Background-Normalized Signa, Contrast Pairs, and Relative Contrast. There are new schedule settings to support these new tests. |
Reference Number | Details |
---|---|
3687 | Fixed issue editing Machine Log incidents when the user was editing an incident where they were not connected to machine the incident was opened under. Prior to the fix, the machine would randomly change to a machine that the user was connected. |
3690 | Corrected issue with excel imports not populating custom numeric tables when there was only a single reading. |
Release 24.10.16
New Features
Reference Number | Details |
---|---|
3627 | Added feature to import cells from an uploaded spreadsheet to be imported into most test fields. The mapping of cells to test fields must first be mapped in QA settings. TG-51 worksheets are excluded from this new functionality. |
3678 | Updated how Decision Support System(DSS) is displayed in reports. Test limit were renamed to action limits, the Summary section will default to expanded, detail sections will default to collapsed, time displayed will be in local time for the account, and the precision in hover graphic was reduced to three decimal places. |
3665 | If a custom numeric table test is modified, the data in that test for any in progress reports (new QA) will be cleared. |
Reference Number | Details |
---|---|
3654 | Fixed issue where titles at the top of the intensity plots for the Catphan wave analysis do not match the correct ramp number. |
3663 | Corrected issue with trends not including data from the current day. |
3675 | Improved registration of MV-QA in the analysis to avoid processing errors. |
3676 | Corrected issue where tasks on the measurement device edit page could not be completed. |
3680 | Resolved Decision Support System(DSS) error message displaying when editing some projects. |
3681 | Fixed issue where Decision Support System(DSS) counting days where energies were not measured. |
3667, 3670, 3679 | Resolved Decision Support System(DSS) issue with users being logged out when attempting to edit a project. |
Release 24.10.03
New Features
Reference Number | Details |
---|---|
3636 | The picket fence analysis has been refined to eliminate artifacts when determining the rotation of the picket fence pattern. |
3580 | Updated radioactive source exchange allow exchange of a source without adding a new source. This handles cases where a machine is being deactivated. |
3651, 3652 | Added analysis options for the Picket Fence analysis to the Image Processing section of the schedule settings. |
3489, 3519 | Added Decision Support System(DSS) analysis feature. This new feature is in Beta testing currently and will be generally available at a future date. |
Reference Number | Details |
---|---|
3644 | Fixed error in finding pickets on very short Picket Fence patterns. |
3635 | Corrected invalid data point displaying on trends accessed from the new QA page in certain cases where there is an open ended tolerance. |
3646 | On TG-51 worksheets where Dose at dref is used, the graphic for the formula was incorrect. This has now been corrected. |
3648 | Refined Picket Fence analysis to detect the correct number of pickets when the pixel size is smaller than 4 mm.. |
3650 | Resolved issue with where the Equipment Hub download button was pointing to the wrong version of Equipment Hub. |
Release 24.09.19
New Features
Reference Number | Details |
---|---|
3619 | Elekta CBCT images may now be manually identified by adding the following ID strings to the Patient ID or Study ID dicom tags of individual slices:
Multiple manual IDs can be set up in a series; however, if any manual IDs are employed, only slices with manual ID strings will be processed. ID strings are case-sensitive. The older ID strings of Contrast, Uniformity, and Spatial will no longer work |
3625, 3626 | Added new X Scaling test for the CT/SIM image processing analysis. |
3608 | Updated Summary Reports page to include an optional section, "Report completed by due date". |
3385 | Updated Create Task page to include the option to create recurring tasks. |
3640 | Updated Machine Log to autofill the machine downtime end with the current time. |
3568 | Changed the rendering of new QA pages so that wide tests have an individual scroll bar rather than a single scroll bar at the bottom of the page. |
Reference Number | Details |
---|---|
3616 | Fixed issue with VMAT open field image was identifying as a strip image. |
3621 | Fixed scaling issue with PRM profile comparison analysis. |
3575 | Resolved issue with tolerances set for energy dependent custom tests not utilizing the "energy unspecified" tolerance settings when there is no preset energy used. |
3634 | Corrected problem where the Reports page presented an error if a user saved a blank report. |
3630, 3631 | Fixed issue with the inability to hide or filter hidden documents within Documents page. |
3642 | Corrected issue rendering special characters in meta-item names for reports. |
3641 | Resolved inconsistency between the test completion indicator for test with NaN results. Both the new QA and report will consider tests as complete if all of the test fields contain some result, NaN or otherwise. |
Release 24.09.05
New Features
Reference Number | Details |
---|---|
3601 | Updated reports page to include a "Finalized Date" column. |
3557, 3607 | Added the ability to create recurring tasks. Also changed the label on the Email Assignee(s) checkbox to Email Assignee(s) on Creation/Update for clarity, made the Email Assignee(s) checkbox sticky to retain its state when a task is reopened, and updated links in notification emails to open the individual task rather than the full task list. |
2942 | Added automatic notifications and banners to alert when Equipment Hubs are nearing end of support. |
3615 | Renamed several CT/SIM Check analysis tests and reorganized the sections.. |
3617 | Added the ability to change source activity date to HDR Monthly test. |
3600 | Updated periodic reports to accept and display filtered date range data points. |
Reference Number | Details |
---|---|
3584 | Fixed issue with custom child tests for daily output results not displaying when attempting to set set default tolerances. |
3587 | Resolved MPC Beam output energies incorrectly being displayed for previous reports. |
3609 | Corrected problem with the output constancy test when removing adjusted rows. |
3613 | Updated Reports page to accept and filter schedules correctly that contain symbols in the name. |
3614 | For the CT/SIM analysis, fixed error in determining the range of slices when looking for the orientation RL markers and added enhancements to make the letters stand out better. |
3620 | Corrected issue with built in system tolerances crashing default tolerance screen. These built in tolerances are currently only in use for the CT IQ analysis for specific accounts. |
3623 | Fixed issue with "Registering files" spinner in new QA displaying for long periods of time when uploading empty folders. |
3624 | Resolved issue where if the user clicks away from “Export Results” on the machine log list page, the contents of the confirmation dialog are still visible on the page |
3629 | Resolved schedule settings issue preventing some tests from being shown/hidden. |
Release 24.07.10
New Features
Reference Number | Details |
---|---|
2805 | Organization name is now displayed below users name for users who work with more than one organization. |
3551 | Updated schedule settings to include the ability to sort tables in reports by meta item. |
3183 | Updated the Starshot test to report angle values for each collimator beam. |
3605 3606 | Added support for Standard Imaging's CT SIM QA phantom. |
Reference Number | Details |
---|---|
3582 | On the Schedules, the most recent report will be shown in black if that report has been voided. |
3586 | Resolved issue with saving the MPC ENERGY (FROM FOLDER NAME) field in QA settings for MPC tests. |
3588 | Updated Picket Fence analysis to account for low resolution image when finding extents of pattern. |
3579 | Resolved issue rendering monthly output test results after they have been marked as NA. |
3603 | For numeric tables, row, column, and table equations no longer require a semi-colon for intermediate calculations. Previously the results for the first calculation in a multiline equation that did not end in a semi-colon would be displayed. |
3604 | Fixed issue where IBA MyQA Daily results were not being automatically pulled when entering new QA. The results will now be pulled upon opening the new QA if the schedule settings are set to automatically pull. |
Release 24.06.26
New Features
Reference Number | Details |
---|---|
3565 | Added the ability to set and change the next due date when finalizing a report. |
3564 | Updated reports to render NaN values when image processing completes but no valid results are generated. |
3532,3533 | Moved Accounts Settings section from the owner's settings to its own page under the Manage page. |
3536 | Added the option to appoint a co-owner in Account Settings page. |
3563 | Updated schedule page to display current status of the latest report. |
3555 | Updated new QA to clear measurements when changing the calibrated equipment selection. |
3573 | Added Collapse All and Expand All buttons to test selection in periodic report. |
Reference Number | Details |
---|---|
3572 | Fixed issue with machine log export results not displaying. |
3540 | Created new error message when there are daily QA device results for the given date but none of those results match the energies mapped to the schedule. |
3574 | Created additional checks for FC-2 phantom to identify validity of center BB detection. |
3576 | Resolved issue with preset tolerances not matching scan description. |
3571 | Improved sign off report functionality to display error message when incorrect password is entered. |
Release 24.06.12
New Features
Reference Number | Details |
---|---|
3342 | For weekly schedules the largest frequency has been increased from 3 weeks to 6 weeks. |
3488 | Updated the API to perform calculations for numeric tables if all needed readings and/or meta-items were uploaded via the API. |
3510 | Changed schedule due lead time used to determine the due date of the next schedule for schedules that used a multiple of the schedule frequency. For example, a monthly schedule set for every three months will now have a 75 day lead time rather than a 25 day lead time. |
3504 | Added sign off due column to report section on Review page to support new option to apply a deadline to report sign off. |
Reference Number | Details |
---|---|
3535 | Updated code library that was interfering with Active Directory login. |
3538 | Improved SNC MV processing thresholding to include dilation to the edge lines when doing the initial thresholding. |
3539 | Corrected issue where setting empty tolerances by meta-item prevented further tolerances by meta-item to be added. |
3549 | Resolved issue where when selecting calibrated equipment besides the schedule default would show as blank in subsequent visits to the new QA in progress. |
3552 | Fixed issue with schedule that are one day overdue, displaying as due today in the Schedule page. |
3550 | Corrected display issues for longitudinal plots for periodic reports when tests were excluded from the report. |
3567 | Added filter for the Machine edit page to disallow certain special characters and to escape the resulting string to avoid potential security issues. |
Release 24.06.05
New Features
Reference Number | Details |
---|---|
3457, 3518 | Added support for PTW Mephysto Beam Profile PDFs. |
3497 | Added system for pre-setting tolerances for built-in tests. |
3499 | Added signoff name to radioactive sources. |
3507 | The lookup page export button is now available from the edit and preview pages. |
3527 | Option to set calibrated equipment w/out default and allow selection in new QA. |
3523 | Improved the speed to load the View Machine Log page upon opening the page or setting filters. |
3509 | For summary reports, meta item and test selection controls inactive (grayed out) unless the schedule text box is checked. |
3524 | Handle a single slice upload for CT591 Slice thickness results |
Reference Number | Details |
---|---|
3508 | Resolved issue with summary reports where the test selection reverted back after deselecting tests and saving. |
3514 | Resolved issue where lookup tables could not be saved. |
3516 | Fixed typo on schedule create and edit pages when daily is the frequency. |
3517 | Corrected report format when "display test instructions" is checked. |
3520 | Resolved issue saving tolerances for a linear regression test where the tolerance for the statistics was also applied to the deviation. |
3525 | Resolved issue with display of tolerance preview graphic. |
3534 | Fixed issue where changing to a different electrometer setting did not clear the output test. |
Release 24.05.23
New Features
Reference Number | Details |
---|---|
3013 | Added the option to set a sign off deadline in schedule edit page and view sign off indications in periodic reports. |
3492 | Added Gamma Cameras to the list of machine types. |
3493 | Updated monthly user account login history to automatically inactivate any user account that has not been used in one year. |
3380, 3496, 3421, 3445, 3428, 3369 | Added support for IBA CT IQ phantoms |
3484 | Added a /GET request to preview how adding results is done through API. |
3500 | Improved the sidebar to include a count of tests completed while in new QA. |
3502 | Improved Isoalign radiation/light field analysis to handle extraneous objects in the field. |
Reference Number | Details |
---|---|
3494 | Fixed cell references when building custom calculations in numeric tables. |
3495 | Corrected issue with Hancock charts not rendering in new QA. |
3501 | Resolved the issue with the top save button on default tolerance page not displaying existing tolerances before the save. |
Release 24.05.09
New Features
Reference Number | Details |
---|---|
3478 | Added option to set deadline (in days) for signing off on reports. |
3375 | Added option to allow new rows to be added to numeric tables in new QA. |
3471 | Updated Sites and Machines page to include option to not allow users selection if they don't have users modify permission. |
3448 | Improved Schedules page to display a Tests Performed section that includes number of tests complete while performing QA. |
Reference Number | Details |
---|---|
3481 | Fixed issue with IBA myQA data clearing preset common data meta items. |
3482 | Corrected issue with incorrect number of tests displaying with In Progress reports after reopening. |
3487 3490 | Resolved issue with added preset rows not saving without requiring data. |
3471 | Updated Sites and Machines page to hide user selection link if they do not have permission to modify users. |
Release 24.04.10
New Features
Reference Number | Details |
---|---|
3458 | Added visualization of tolerances to the edit tolerance page. |
3459, 3461 | Added the option to add a reference image for Dose lab kV and MV analysis. |
3394 | Released new Equipment Hub version 3.4.3.0 which added support for IBA myQA Daily device. |
2954 | Added option to hide rows for numeric tables in schedule settings. |
3470 | Added the ability to sign-off on periodic reports. |
3473 | Added the ability to select and/or deselect individual tests in periodic reports. |
3480 | Updated Somatom Analysis to accept alternate titles for processing. |
3474 | Created variables for PTW Mephysto beam profile. |
Reference Number | Details |
---|---|
3463 | Fixed issue with Longitudinal screen not rendering correctly for numeric variables. |
3464 | Corrected issue with custom numeric tables where some row names were not accepted. |
3465 | Fixed issue where hidden tests were included in complete/incomplete total in reports. |
3370 | Resolved incorrect display of tests completed and tests performed in finalized QA and Reports page. |
3466 | Fixed Longitudinal trends missing from new periodic reports. |
3467 | Updated IC Profiler PRM tests to accommodate when both collimator and gantry angles were included in the PRM file. |
3472 | Resolved meta items not rendering preset values in New QA for lookup tables. |
3476 | Improved algorithm for finding MTF bead for the Elekta Catphan analysis. |
3477 | Improved thresholding to allow poor quality Elekta Catphan images to process normally. |
Release 24.03.28
New Features
Reference Number | Details |
---|---|
3442 | Machine Log notifications to vendors will now include the incident number. |
3333 | Updated Beam Checker QA to allow selection/deselection of individual tests in the schedule settings. |
3452 | The edit page for measurement devices will now display time stamps in user's timezone. |
3403 | Periodic reports are now granular down to the individual test level. |
3460 | Updated task subject line in email notifications for due tasks to read "Your task is soon to be due!" |
Reference Number | Details |
---|---|
3447 | Corrected units for MPC Rotation Offset test to degrees. |
3433 | Fixed issue with inconsistency between new QA completed tests displayed and the number of completed tests displayed as completed on reports page. |
3450 | Corrected error in estimated rotation on Ethos Picket Fence analysis. |
3454 | Updated displayed measurement units for max position error for MPC Couch test to mm. |
3453 | For the Ethos Picket Fence analysis, the Leaf Deviation Summary column was renamed to Median Deviation to avoid confusion. |
Release 24.03.07
New Features
Reference Number | Details |
---|---|
3262 | Updated Report page with more options to filter by schedules, sites and machines. |
3440 | Added Energy meta items for FS-QA phantom analysis for cases where the energy can be read from the DICOM tags. |
Reference Number | Details |
---|---|
3434 | Corrected issue with numeric tables where tables with more than one row could not be updated via the API. |
3436 | Fixed issue where expected source activity for brachytherapy could be prefilled from the browser cache and override the calculation. The browser cache will now be ignored for source activity. |
3438 | Resolved issue where tolerances imported from a daily QA device were shown in new QA when unspecified energy tolerances were set. The report contained the correct tolerances. When any tolerance is set it will override the tolerances from the daily QA device. |
Release 24.01.31
New Features
Reference Number | Details |
---|---|
3408 | Added unspecified energy to tolerances for energy dependent tests. |
2455 | Improved edit tolerances with the options to set a single tolerance for any meta item and an option to set a single tolerance for applicable energies. |
3406 | Updated Catphan analysis to include Noise Power Spectrum tests. |
3392 | Updated date options for Summary Reports to include "Last 7 Days" and "Previous Month". |
3416 | Added analysis type to Image Processing emails. |
3419 | Improved application of schedule settings to reset in progress tests and apply any new settings, if schedule settings are changed. |
3425 | Updated custom tests for angular deviation to allow nominal and difference fields to have units other than degrees. |
Reference Number | Details |
---|---|
3409 | Fixed issue with the inability to copy custom tests that included lookup table meta items. |
3414 | Corrected calculation issue with lookup table meta items displaying incorrectly in new QA. |
3418 | Resolved FSQA analysis crashing when Y direction leakage is very high. |
3422 | Improved thresholding with FC2-kV Scaling images with significant shading. |
3423 | Fixed an incorrect variable reference with Picket Fence analysis when thresholding fails. |
3426 | Corrected issue with the inability to download templates from Template Library. |
3424 | Resolved custom test issue with the inability to store pass fail labels with < or > symbols. |
3429 | Fixed ACR CT analysis situation where a large phantom tilt caused extra slices to be cut off. |
3431 | Corrected issue when choosing "Use new default tolerances" would clear all tolerances from schedule. |
Release 24.01.17
New Features
Reference Number | Details |
---|---|
3267, 3384 | Added support for Quart DVTvn model phantoms with the additional water plug in sensitometry module. |
3388 | Updated the star shot analysis using TIFF images to accept meter as an acceptable resolution unit. |
3357 | Improved report list screen to update in progress reports every time that report is saved. This allows the saved report to display the most recent incomplete and complete numbers. |
3396 | Updated custom test lookup tables to allow multiple fields for single meta item. |
3399 | Added the ability to display all available slice thickness measurements on Catphan results. |
Reference Number | Details |
---|---|
3381 | Fixed situation in Elekta CAT where all raw plug values displayed error when calculating sensitometry linearity. |
3382 | Corrected issue with Caphan analysis where .mat files uploaded for a different analysis were causing Catphan image sets not to process. |
3368 | Resolved issue with custom test section instructions not showing up in new QA. |
3391 | Fixed issue with the inability to delete lookup tables. |
3395 | Corrected issue with saved date meta items not rendering in new QA. |
3405 | For Catphan image sets, the HU value threshold needed to identify the CTP 515 low contrast module was relaxed. |
3404 | Fixed issue where the series descriptions in new QA still display for removed image sets. |