Image Owl uses the continuous integration and continuous delivery (CI/CD) development model to ensure that we release new changes to our customers quickly in a sustainable way. We deploy small batches of code changes to production regularly, usually weekly so that the code is fresh in developer's mind and there is a limited amount of dependencies to test. This allows us to be responsive to customer needs for bug fixes and new features while maintaining a robust and well tested code base.
Note: Equipment Hub versions are documented in the Revision History article. That revision history only contains changes to the Equipment Hub service itself. There may be related changes that are listed below that may also be of interest.
Release 25.12.04
Release Date: 12/30/2025
New Features
| Reference Number | Details |
|---|---|
| 4238, 4246 | Added support for performing the Asymmetric field test on Iso Align and FSQA radiation/light field coincidence phantoms. |
| 4190, 4212 | For Corgi phantoms, added the ability to download the raw noise power spectrum data into a comma separated value (CSV) format. |
| 4187, 4188 | Added an optional test to be added that provides a diagnostic log on radiotherapy based image processing. |
| 4235 | Created new end point for the Equipment Hub File Uploader for organizations that block https://s3.us-east-1.amazonaws.com. The new optional endpoint is https://tqa-imageowl-upload.s3.us-east-1.amazonaws.com. Please contact support to move to the new address. |
| 4202 | When using individual image processing widgets it was easy to miss image sets that had not been processed. New QA cannot be finalized until all image sets are processed. A new popup dialog has been added when attempting to finalize a report that lists the which analysis have image sets awaiting processing. |
| 4170 | Updated the API get call for Machine Logs to accommodate custom fields and fault captures. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4245 | Fixed a Catphan analysis issue where misidentification of a slice in the CTP682 module led to a failure for Catphan model 605 phantoms. |
| 4230 | Fixed a Test Settings error where default settings could not be set for the Catphan analysis. |
| 4233 | For Corgi, fixed an issue on rendering reports when one of the Corgi modules could not be analyzed. |
| 4198 | For Test Settings, changed the behavior when setting global tolerances where the baseline was left blank. Prior to the change a blank baseline could be applied over a schedule that had a baseline set. Now blank baselines will only be applied to new schedules that do not have a baseline set. This does change how absolute tolerances are set. |
| 4247 | Resolved issue with image processing starting before all files are uploaded if the account was using the https://tqa-imageowl-upload.s3.us-east-1.amazonaws.com endpoint from the fix for 4235 listed in the new features for release 25.12.04. |
Release 25.11.10
Release Date: 12/9/2025
New Features
| Reference Number | Details |
|---|---|
| 4072 | Allow editing of TG-51 worksheet test descriptions, settings, and tolerances from the Manage - Test Settings page. |
| 4003 | For Corgi, added new 45 degree MTF tests. |
| 4196, 4197 | For IC Profiler, added schedule setting for PRM file imports, that when selected, will try to discern the beam type and value from the filename field if the energy fields are set to undefined. |
| 4194 | For IC Profiler, added support for measuring Diagonal Flatness on PRM file imports. |
| 4157 | In preparation for a future beta test, the ability to add communication groups has been added. The Manage - Communications Groups will only be available after being enabled by support. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4210 | For Tomophan, fixed an issue with noisier Tomophan images that could orient the images incorrectly, top to bottom. |
| 4205 | Corrected issue with tolerance history date/user not updated for current tolerances. |
| 4215 | Fixed issue for custom numeric tables where calculations using an assignment were not displayed in the new QA page but were visible in the report. |
| 4214 | In the Image Processing setting section of the schedule settings, settings can be reset to the default settings on a per analysis basis. |
| 4217 | Fixed an issue where scan dates on in new QA pages were blank when calculation notes were added. |
| 4135 | Resolved issue where uploading file to a custom file upload test with children tests would mark the children tests as "undone" on report's timeline. |
Release 25.10.27
Release Date: 11/18/2025
New Features
| Reference Number | Details |
|---|---|
| 4148 | When a user reorders a section in the Review page the section will now be reloaded rather than the full page. |
| 4169 | Fault Captures designated as updates to previous fault captures will now display full details of the fault. |
| 4072 | Editing of TG-51 test descriptions, settings and tolerances is now supported in Manage - Test Settings. |
| 4144, 4145 | Added analysis for the Varian RapidArc Dynamic Collimator Speed test. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4162 | For the kV-QA phantom, the thresholding parameters were adjusted to support Hypersight images. |
| 4179 | Fixed Custom Numeric Table issue where the custom fields for measurement devices could not be added in an Equipment type table meta-item. |
| 4164 | Fixed the display of TG-51 tests on the machine level Trends page. |
| 4171 | Fixed an issue with retaining DICOM tag filter operators when copying File Uploaders. |
| 4091 | Eliminated a spurious second run for certain scheduled periodic report configurations. |
| 4180 | Fixed an issue preventing the saving of periodic report configurations. |
| 4153 | Corrected issue with History Button within Test settings displaying an error. |
| 4184 | Fixed an issue in Test Settings where the "Set all..." buttons were not functional when saving baseline tolerances for energy dependent tests. |
| 4185 | Fixed custom Numeric Table issue where a lower precision was used when previewing the table versus the same Numeric Table in a schedule. |
Release 25.10.22
Release Date: 11/04/2025
Bug Fixes
| Reference Number | Details |
|---|---|
| 4149 | Corrected issue when entering a new password with special characters, where the password was seen as different between the two entries. |
| 4155 | Resolved issue with default descriptions on variables from being overwritten with account specific descriptions. |
| 4161 | Fixed an issue in schedule settings where presetting energy tests with multiples of the same energy displayed only the energy once. |
| 4163 | Fixed an issue when setting tolerance by meta-item where the value of the meta-item was hidden. |
| 4168 | Corrected issue when attempting to save blank tolerances where an uncaught exception was displayed. |
| 4138 | Updated Account Settings and User Account to confirm the owner role can only be assigned by the account owner or support. |
| 4156 | Fixed a spelling mistake and inconsistent punctuation on the 4D CT Analysis. |
Release 25.09.17
Release Date: 10/23/2025
New Features
| Reference Number | Details |
|---|---|
| 4045 | File Uploader filters now allow OR operations when filtering on a specific DICOM tag. Any tag that is added more than once will perform an OR operation on the values in all identical tags. |
| 4103 | The File Uploader will now separate CT and MR image sets by the Series UID (tag 0020,000e) uploaded through a File Uploader filter for separate processing. |
| 4108 | Machine Log visibility is now editable under the Manage - User Roles page. |
| 4053 | Improved MPC PDF Import handling of unusual spacing in the PDF file. |
| 4069 | Updated filters for calendar section on Review page to remember the last setting and use that setting when returning to the page. This functionality uses cookies and is limited to a browser instance. The setting will be lost when clearing cookies on the browser. |
| 4104 | Added ability to create custom text, list, or yes/no/NA fields for machines under Manage - Sites and Machines. The fields can be available for all machine types or if desired for specific machine types. |
| 3923 | Added protection for historical reports to make sure they don't change over time when tests are changed. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4033 | Corrected issues when creating charts where either a error was seen or with text overlapping the graph when navigating to trends and using the back button to return to the report. |
| 4129 | Fixed formatting issue on bar graph x-axis labeling after chart library upgrade. |
| 4147 | Fixed the issue where chart scaling was far too large after chart library upgrade. |
Release 25.09.03
Release Date: 10/9/2025
New Features
| Reference Number | Details |
|---|---|
| 4096, 4099 | For the Leeds analysis, added Uniformity, Contrast, Contrast to Noise Ratio, and Noise tests using the central uniform objects. |
| 4077 | For Doselab MV phantoms the segmentation parameters were adjusted to meet a wider range of images which will improve ROI placement. |
| 4116 | Improved Catphan analysis to handle slice location (0020,1041) tags that do not contain a value. |
| 4115 | Added monitoring code to new QA pages to produce more fine-grained internal logging. |
| 4121 | Updated charts library to improve charts shown in QA reports, summary reports, and trends. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4089, 4094 | Fixed an issue in accessing Test Settings for some Catphan tests. |
| 4090 | Corrected issue with spurious error and popups on TG-51 Schedules when temperature and pressure were not filled out. |
| 4080 | When using the API report comments can now be added to an in progress report.. |
| 4095 | Resolved issue when editing Equipment Hubs where measurement devices are unchecked during an update. |
| 4100 | Corrected issue for single sign on users where the inactivity time out was set to 6 minutes after making any edits on the Accounts Setting page. |
| 4110 | Fixed issue on the Magphan RT where it failed to find the 1 cm spheres on certain image sets. |
| 4066 | Fixed issue on the Review page where columns could not be sorted. |
| 4111 | Fixed an API issue where the number of files found when activating File Uploader was under reported. |
| 4091 | Resolved an issue where periodic reports scheduled to run on a specific day of the week were run twice. |
| 4113 | For picked fence, the thresholding parameters were adjusted to avoid detecting spurious junctions in specific sets. |
| 4119 | Inactive schedules are no longer shown in the schedules drop-down for the calendar on the Review page. |
Release 25.08.20
Release Date: 9/9/2025
New Features
| Reference Number | Details |
|---|---|
| 3975 | Changed the default wildcard for fault capture from *.png to Print*.png. |
| 3980 | Added the ability to list files matching the Fault Capture wildcard in the watched folder. |
| 3981 | Added the ability to see logged Fault Capture events in the Fault Capture configuration page |
| 3983 | Added a button to display the Fault Capture processing log for a selected Fault Capture. |
| 4070 | Adjusted ROI the constraints for MV Imaging (MV-QA) to avoid "Processing time has exceeded the maximum allowable time of 5 minutes" error. |
| 4058 | Changed to a new input handling library to better handle < and > characters where needed but reject HTML tags for security. |
| 4064 | Updated Manage/Users screen to only allow support and owner role to have the ability to edit the owner users role. |
| 4078 | Users will now be required to provide their current password before changing their password or email in their account settings. |
| 4079 | Updated custom device fields to disallow text that could be interpreted as HTML tags. |
| 3890 | Improved Las Vegas analysis to support smaller unity field than 30X30. |
| 4081 | Added Referrer-Policy and Permissions-Policy headers to all hosted web sites. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4061 | Fixed issue where report page could crash if a test had a tolerance set but no calibration factor was linked to the energy. |
| 3963 | Resolved issue with fault capture date settings not capturing earlier set dates. |
| 4057 | Fixed issues where a checkbox custom test would display as an action limit exceeded in reports. |
| 4071 | Corrected issue with SNC MV-QA analysis center pixel not calculating. |
| 4044 | Resolved issue with File Uploader failing on files that were over 6 MB. |
| 4075 | Resolved issue triggering the File Uploader in a schedule via an API call. |
| 4073 | Corrected issue with data export failing with an error status. |
| 4074 | Fixed an issue leading to a the misidentification of a CTP700 wave module as a sensitometry module. |
| 4041 | Fixed issue with numeric tables where calculations between two dates returned the number of seconds between the two dates rather than the resultant date. |
Release 25.08.06
Release Date: 8/21/2025
New Features
| Reference Number | Details |
|---|---|
4048 |
Upgraded pdf.js library to latest version. This should have no effect on the user experience. |
4055 |
When image sets are uploaded via the API, we will now automatically include any module files that have been added to the machine. It is no longer necessary to include the module file in the upload. |
Bug Fixes
| Reference Number | Details |
|---|---|
4046 |
Resolved issue where comments were not properly rendered in a report when files were uploaded to an image processing test between the time the comment was created and the new QA was saved. |
4056 |
Resolved Catphan error handling that caused a processing failure when calculating position and orientation statistics |
Release 25.07.23
Release Date: 8/12/2025
New Features
| Reference Number | Details |
|---|---|
| 4021, 4022 | Added MLC naming convention for TIFF starshot images. Starshots that use mlc in the name will be categorized separately since the current naming convention includes gantry, table, and collimator but the collimator designation does not indicate whether the MLC or jaws were used. |
| 4015 | Adjusted thresholding for Doselab kV so that a wider array of images will be supported. |
| 4036 | The thresholding for QC-3 and QC kV images has been better optimized for a wider range of acquisitions. |
| 4006 | Added a history button for test description that allows users to see prior changes and who made them. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 4026 | Resolved issue under Manage - User Roles where when modifying a user role the schedule selection was not being saved when the select all button was used. |
| 3990 | Fixed issue where only the built in role would display on the Manage Users page if a custom user role was created with the same name as a built in user role. |
| 4018 | Corrected an issue where a user could see a test description outside of their account if they manually changed the test number in the web browser address window. |
| 4035 | Fixed incorrect units on the HU Constancy test in the 4D CT Analysis. The test now displays HU as the units. |
| 4029 | Corrected issue where settings could not be saved for the X-Ray Output Constancy test when the account was set not to allow data comparisons. |
Release 25.07.02
Release Date: 7/25/2025
New Features
| Reference Number | Details |
|---|---|
| 3967 | Added refinements to SNC MV registration when taken at 2.5 MV. |
| 3984 | Added refinements to SNC kV Registration for Ethos kV-QA Phantom ROIs Offset. |
| 3982 | Added the ability to easily filter to view only Fault Captures in the Machine Log listing. |
| 3989 | Updated login sessions to render a message to users if their session has ended and they need to login again. |
| 4001 | FS-QA phantom energy meta-items can now be retrieved from multiple tags in the DICOM headers to handle cases where the energy value is not found in the RTImageDescription tag. |
| 3744 | Updated the graphic incomplete plots in new QA and schedule settings to avoid confusion with file attachment tests. |
| 3912 | Updated users roles to provide a history of changes made. |
4007 |
Added a "usage" button to the individual tests on the Manage Custom Tests page that display templates and schedules that use the test. |
4005 |
Adjusted parameters on Catphan slice thickness algorithm to reduce the likelihood of reverting to the previous algorithm. |
3946 |
Improved load times for calendar view on the Review page when changing the time period. |
4010 |
Now providing a more informative error message when strips in a Hancock test are touching the top or bottom edges of the image. |
3924 |
If Equipment Hub installed worked but not connection was made afterwards, the Equipment Hub still shows up as installed in the system. An error message and possible solution is found in the Log for the Equipment Hub |
3910 |
Enhancements made to the Fault Capture auto-close options. |
4016 |
Improved thresholding on high-resolution, low mAs SNC kV phantom images. |
4017 |
For the starshot analysis, the keyword "couch" may be used in TIFF starshot images in addition to table to direct the results into the table starshot tests. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3999 | Fixed Hancock analysis issue with handling non-monotonic leaf profiles. |
| 4014 | Corrected issue when uploading a file to a binary test in an empty QA causes "QA Saved Elsewhere" popup message. |
Release 25.06.18
Release Date: July 8, 2025
New Features
| Reference Number | Details |
|---|---|
| 3987 | Added support for import of Somatom Confidence Daily PDF's. |
| 3953 | When reverting back to defaults on a custom tests, added a popup that confirms whether or not the user wants to reset the test instructions. |
| 3905 | For custom test meta-items, added option to sort the meta-items, edit the units, or delete them. When deleting a meta-item any schedules that have presets for that meta-item will be shown. |
| 3968 | Added the option to set tolerances for variables in 4D CT Analysis. |
| 3878 | Improved handling of data save conflicts when a user is logged into the same new QA in more than one tab or browser. These instance often result in blank tests being saved over completed tests. If a save or finalize is attempted that would conflict with the prior save in another location, the user will be prompted to record the data in the current session. In cases where the conflict is a prior save, the user will be further prompted to refresh the page and reenter any needed data. In cases where there is a prior finalize the user will be further prompted to go the the report and reopen the report if needed and if they have permission. |
| 3992 | Improved segmentation on lower resolution for Doselab MV images. |
| 3973 | Improved handling of inverted images for failing Doselab kV Images. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3956 | Fixed issue selecting baseline from a finished report where the green highlight for baselines was mispositioned. |
| 3971 | Fixed issue where trend results could not be pulled for IC Profiler PRM tests via the API. |
| 3972. 3974 | Fixed issue in schedule settings when setting the next due date, where the displayed date was shown according to UTC rather than the local timezone. |
| 3970 | For Catphan phantoms, improved the algorithm that determines which slices are valid slices. |
| 3988 | Doselab Planar MV images uploaded as a set will now yield results for any images that can be processed. Prior to the change all images in the set would be rejected if one or more images in the set failed. |
Release 25.06.04
Release Date: June 24, 2025
New Features
| Reference Number | Details |
|---|---|
| 3945 | The Review page header is now frozen in place when scrolling so that collapse and expand buttons are always visible. |
| 3941, 3940, 3939, 3901, 3902, 3903, 3904, 3952, 3364 | Added 4D CT analysis for analyzing multiple Catphan scans by phase. |
| 3947, 3948 | Added support for alternative ways of determining identity of Hancock analysis. This includes adding meta items for Gantry, Table and Collimator. The gantry, table and collimator angles are displayed in the report if available. |
| 3930 | When creating new roles there is now an option to select schedules that the role should have access to. |
| 3819 | Final report comments have been updated to display time stamps. |
| 3962 | Added security headers to all HTTP requests to enhance security. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3943 | Added a refinement to Sun Nuclear MV-QA phantom analysis to handle inverted images. |
| 3955 | Fixed issue in schedules settings with importing spreadsheets where user could not manually enter cell numbers. |
Release 25.05.21
Release Date: June 5, 2025
New Features
| Reference Number | Details |
|---|---|
| 3866 | When managing sites and machines, the edit machine page will now display associated tasks and documents. |
| 3863 | For numeric tables users can now add numeric and text fields in addition to readings. Tolerances may be set on numeric fields and numeric fields may be used in column, row, and table calculations. |
| 3813 | The Review page now includes a schedule specific calendar that displays the schedule status. |
| 3785 | When editing tests in the Test Settings page, the test name will be shown at the top . When exiting the edit page the user will be returned to their original spot in the Test Settings page with any sections expanded or collapsed as they were prior to opening the edit page. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3934 | Fixed issue where the chosen template name is not populating in the name field when adding a schedule to a machine. |
| 3933 | Corrected Catphan issue where a Catphan 600 could be misidentified as CT504 Phantom. |
| 3935 | Resolved issue with the inability to save tolerances from full tolerance page when there were meta-items with the same name in different tests. |
| 3938 | Resolved issue saving documents where the associated machine was not saved in the document configuration. |
Release 25.05.08
Release Date: May 22, 2025
New Features
| Reference Number | Details |
|---|---|
| 3894 | Updated Task due soon email notification to include due date in subject line. |
| 3862 | Updated report timeline to include removed image processing uploads. |
| 3891 | Improved Las Vegas analysis to include alternative thresholding method. |
| 3830 | Refactored scheduling code so that schedules, DSS Projects, periodic reports, and tasks are handled using the same code set. |
| 3897, 3899 |
For IC Profile added the following tests for PRM import:
|
| 3898 | Added error message for Fault Capture when a redacted folder check is performed but the Equipment Hub version is to low to support the check. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3893 | For FSQA Scaling, improved the handling of ROIs with uneven background to better locate arrows. |
| 3892 | Fixed erroneous incomplete test warning in finalize dialogue. |
| 3871 | Fixed issue where users could set multiple tolerances by meta-item using the same value for the meta-item (i.e. 5 and 5.0). When matching the meta-item we will use the first matching value. After the fix, users may not set two meta-items that evaluate to the same number. |
| 3882, 3885 |
Made several improvements for the Corgi phantom:
|
Release 25.04.16
Release Date: May 8, 2025
New Features
| Reference Number | Details |
|---|---|
| 3837 |
Added a scan date and time column to the to the upload information in new QA and reports. For DICOM files this will use the tags to determine the scan date and time; we will assume that time is in the local time zone. We will search the following tags in the order shown below, the first full set found will be used:
For all other files we will use the upload time. |
| 3839 | Updated custom numeric tables to allow saving while editing without redirecting from screen. There is now a Save button and a Save & Exit button. |
| 3875 | Updated report timeline to mark a user as the creator of a report when they are the first to press save or run a processing job. |
| 3861 | Updated report viewer timeline with a View link for each item so that results for each specific change may be reviewed. |
| 3881 3887 | For the Hancock Analysis, added new meta-items to the rotation-corrected results so that show how much the images were rotated. |
| 3877, 3879 | Added scaling image analysis test for the Sun Nuclear FS-QA phantom. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3886 | Resolved issue where the report timeline rendered the report initialized date showing the report date. |
| 3860 | Fixed issue with tolerances where non-numeric values could be added to a numeric meta-item. |
| 3869 | Corrected issue where schedules could not be copied if the "last day of month" was selected or the schedule to be copied was in a schedule group. |
| 3872 | Resolved issue where Decision Support System (DSS) notification emails were not sent. |
| 3858 | Corrected issue with Winston Lutz image analysis where centers could not be found when using very small targets. |
| 3880 | Fault Capture will now poll the specified folder rather than relying on system file creation event. In most environments the file creation event was not triggered. |
| 3884 | Fixed issue where the Manage - Radioactive Sources page rendered source calibration dates in UTC rather than the local time zone. |
Release 25.04.04
Release Date: April 22, 2025
New Features
| Reference Number | Details |
|---|---|
| 3843 | Updated the numeric table builder to display all cell IDs when building custom equations. Cells that cannot be used will have their IDs greyed out. |
| 3848 | Updated Catphan 600 analysis to accommodate new spatial all-air spatial rods used in newer phantoms. |
| 3849 | The CTP 600 preset template has been changed to include the Critical frequency wire test and to remove the obsolete Critical frequency wire [cycles/cm] test. |
| 3836 | Improved numeric tables with the ability to change the order of meta items, calculation rows and columns. |
| 3856 | Improved the Picket Fence analysis to handle bad pixels in the corners and edges for Elekta images. |
| 3852 | Added support for MPC pdf import using file uploader. |
| 3826 |
Updated reports to show a timeline of the following:
|
Bug Fixes
| Reference Number | Details |
|---|---|
| 3838 | Corrected numeric table issue where meta items were unable to be deleted once they were preset in schedule settings. |
| 3828 | Fixed inconsistent reporting of Catphan results, now all contrast detail tests will report NaN when there are no results rather than a blank. |
| 3851 | Resolved issue with MPC tests no longer populating tolerances for test that are newly added to the template. |
| 3855 | Fixed issue with the Catphan analysis to avoid a processing failure for Catphan 500 phantoms. |
| 3857 | Removed extraneous whitespace area when viewing reports on mobile phone. |
| 3804 | Fixed issue where empty custom tests groups could not be deleted. |
Release 25.03.12
Release Date: April 7, 2025
New Features
| Reference Number | Details |
|---|---|
| 3787 | Updated account settings to allow owners and co-owners to control login settings. |
| 3822 3824 | Updated Elekta CAT analysis to force processing when ‘Elekta CBCT CAT Test' is selected in cases where the manufacturer tag is not Elekta. |
| 3816 | Updated the TotalQA logo. |
| 3664 | Refactored signoff query code in view report screen for faster rendering when opening reports. |
| 3774 | Improved the tasks display on measurement device pages to include tables that include upcoming tasks, completed tasks, and documents. |
| 3831 | Added a more informative error message for Winston Lutz image processing if angles cannot be determined from the DICOM tags for filename. |
| 3750, 3764 | Added option to import Varian MPC check report and session report PDF files. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3818 | Fixed issue with upload widget where clicking the "Start upload" button would result in an endless loop of registering files. |
| 3817 | Corrected issue editing periodic reports where numeric table calculations could not be properly selected. |
| 3825 | Resolved empty machines not appearing on Manage - Schedules Configuration page. |
| 3827 | Fixed issue in numeric table calculations where using the source activity widget as a variable in a calculation used the date rather than the source strength. |
| 3829 | Updated Corgi analysis to more reliably produce MTF results. If sigma is less than the pixel size, it will be set to the pixel size when calculating the limits to crop the LSF. |
| 3820 | Corrected meta-Items in numeric table displaying out of order. |
| 3840 | Fixed issue with radioactive source certificate time being displayed in UTC rather than the local time zone in the update history. |
Release 25.03.05
Release Date: March 20, 2025
New Features
| Reference Number | Details |
|---|---|
| 3772 | Added a "+ Schedule Group" on the Manage - Schedule Configurations page to allow the creation of schedule groups where schedules may be grouped and that group may be expanded or collapsed. |
| 3795 | Updated the printing and exporting of reports so that the site, machine, and schedule are shown at the top of the export or print. |
| 3798 | Added the capability to view and find selected tests in the template builder. The magnifying glass icon in the Selected test pane will show that specific test in the Available tests pane while the eye icon will show a preview of that test. |
| 3747 | Added the a date meta-item for custom numeric table tests. |
| 3815 | Made the reference files for SN Planar kV Imaging (kV-QA) and Planar MV Imaging (MV-QA) downloadable from QA settings. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3803 | The GE CT QA phantom analysis is now better able to threshold noisy images in order to correctly locate the contrast module. |
| 3766 | Fixed schedule settings with monthly frequency to correctly use the last day of the month. Prior to the change, when a schedule is set to run monthly on day 31, 30 or 29 and the following month has fewer days, the next occurrence date is moved into the following month (i.e. 31st of Jan → 3rd of March). |
| 3796 | Resolved issue with periodic report email link not containing same test filter as original which resulted in more tests than selected being shown. |
| 3806 | Corrected a Catphan issue with calculating slice thickness where the profile of wire ramps was picking up other test objects. The profile now is limited to avoid picking up other targets. |
| 3807 | Resolved issue with custom numeric test not showing results in the Analysis page when the when the numeric table was edited and one of the original calculations was removed. |
| 3811 | Fixed issue with MPC tolerances not being displayed for MPC tests that were added after the initial MPC import was configured. |
| 3814 | Corrected issue with reopen button option not displaying in some cases after a report was unsigned by support. |
Release 25.02.12
Release Date: March 3, 2025
New Features
| Reference Number | Details |
|---|---|
| 3783 | Improved the loading speed of reports with a large number of images. |
| 3733 | Added checkbox meta item option for regular custom tests and numeric tables. |
| 3789, 3748 | Created pdf import for QUASAR(TM) MRgRT Insight analysis. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3780 | Corrected an issue where processing tests preset as NA rendered as NA in the report despite the N/A box being cleared in new QA. |
| 3784, 3790 | For the Hancock analysis, improved the edge finding algorithm on strip extents and detection of strip extents on faint strips. |
| 3786 | Fixed issue where changing account test settings removed account tests settings from other accounts using the same test. |
| 3794 | Added a check on CTP600 analysis to determine if any module identifications may have been missed. |
| 3799 | Corrected an MPC import issue when the import was triggered via the API and the new QA was subsequently edited where the results did not match the energies. |
Release 25.01.29
Release Date: February 18, 2025
New Features
| Reference Number | Details |
|---|---|
| 3709 | Updated date picker widget to keep from reverting to last set date when only changing the month or year. |
| 3763 |
The CT/SIM analysis has been refined to improve the accuracy of the following:
|
| 3743 | Added a "Day of" option to task reminders that will send a reminder the day the task is due. |
| 3769 | Improved MPC import energy results to only generate processing results for tests within the schedule. |
| 3762 |
Created the following MPC Import variables:
|
| 3754 | Added a real-time equipment hub logging system 3.5.0.0 which added support for Machine Log fault capturing. |
| 3775, 3685, 3711, 3712 | Updated machine log process to include Fault Capturing per account. |
| 3618 | Included site, machine and schedule to error messages when another user is in a QA. |
| 3771 | For the Catphan analysis, improved the accuracy of detecting the z-center of the uniformity and the clarity of orientation images. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3760 | Fixed issue with machine log downtime incorrectly being calculated where machine downtime is calculated over multiple days. Prior to the fix 1 second was being lost for each day. |
| 3768 | Corrected the task email notification so that the links point to the correct page. |
| 3765 | Resolved machine log issue with date time picker rendering behind other elements. |
| 3633 | Fixed Equipment Hub installer error displaying wrong minimum .NET Version. |
| 3770 | Corrected Incident Learning System (ILS) role appearing in accounts as an option that don't have ILS. |
| 3767 | Resolved issue where periodic report emails were not sent for some frequencies. |
Release 25.01.08
Release Date: January 28, 2025
New Features
| Reference Number | Details |
|---|---|
| 3745 | Updated edit template page to include a list of all schedules using the template. |
| 3741 | Added log of changes made to template page. To see the logs click the Historical Changes button in the upper right. |
| 3751 | Added log of changes made to Calibration Equipment section within schedule settings page. To see the logs click the Historical Changes button in the upper right. |
Bug Fixes
| Reference Number | Details |
|---|---|
| 3739 | Fixed an issue within DSS setup where child tests unable to be selected. |
| 3749 | Corrected an issue with MPC Import where tolerances were not displayed in the report for MPC tests that were added after the initial MPC setup. |
| 3753 | Resolved issue with the Electron Dose Rate Output Constancy not calculating output. |