Errors for Equipment Hub will appear on the lower right corner of the browser.
The error will only appear when trying to test the connection when creating or editing the measurement device under Manage Measurement Devices.
Below is a list of common fixes for the "Error while looking up trend data for PTW Quickcheck. Could not find a part of the path '[given_path]." error.
Ensure that the database path is correct. |
Ensure that the database path is correct and that the user the Equipment Hub is logged in as has permissions to reach the file. |
Ensure that the database server name or IP address is correct if the database is not on the same computer as the Equipment Hub. Use only the DNS name or IP address. Do not precede the name with '\\'. |