Errors for Equipment Hub will appear on the lower right corner of the browser.
The error will only appear when trying to test the connection when creating or editing the measurement device during new QA.
Below is a list of common fixes for the "Error while looking up MPC Import files. Could not find a part of the path '[given_path]'." error.
Ensure that the last 4 digit of the linac serial number is correct in the schedule settings. |
Ensure that the path is correct and that the user the Equipment Hub is logged in as has permissions to reach the file. |
Ensure that the server name or IP address is correct. Use only the DNS name or IP address. Do not precede the name with '\\'. |