Prerequisites
Before attempting to configure the StarTrack3 import, ensure that the following items are in place:
- the latest Equipment Hub version is installed
- the IBA StarTrack tests must be included in the template used for the schedule where the measurements will be captured (this is automatic for myQA StarTrack3 branded accounts)
- the location of the StarTrack3 database
- a service account or user who has file access to the StarTrack3 database
Creating the Measurement Device
To create a StarTrack3 navigate to Manage - Measurement Devices, then click the +Add Device button.
The Add Measurement Device page lets you update a device's type, label, site, status, serial number, Equipment Hub and database connections, and manage its documents and tasks. Changes take effect when you click Sign and Update device.
1. The fields show will change based on the Device Type. Select StarTrack from the dropdown.
The left column holds the device information while the right column lists the device's Documents, Upcoming Tasks and Completed Tasks. Back returns to the device list, and View Historical Changes shows the device's change history.
Device Details
| Field | Required | What it does |
|---|---|---|
| Device Type | Yes | The kind of device (for example StarTrack, Ion Chamber, Electrometer, PTW Quickcheck). Choose Add new... to create a custom type. |
| Device Label | Yes | The name shown throughout the application and in the page title. |
| Site | No | The site the device belongs to. |
| Device Location | No | Free text, such as the room or cabinet where it is stored. |
| Status | Yes | Active devices are available for QA; Inactive devices are retired from use. |
| Device Description | No | Free-text notes about the device. |
| Device Picture | No | An image of the device, shown at 150 × 150 pixels. |
| Serial Number | No | The manufacturer's serial number. |
Fill in all of the required fields:
- Device Label - a name is required, it will become the name of the button used in the new QA so it is best to use something users will recognize.
- Site - the site is optional but if one is chosen, the device will only be available for selection on schedules for that specific site. Leave this blank to use the device on more than one site.
- Status - the device must me active to be used in a schedule
Equipment Hub
At least one Equipment Hub must be selected for the device to be used. The Equipment Hub dropdown links the device to one or more hubs. Click x to remove a hub, or + Add Equipment Hub to link another. The Equipment Hub must be on the same host as the database or logged in as a user that has permission to access the database on another host.
Database Connections
There are two databases that must be connected: RAW DATA and MEASUREMENT DATA. Each has its own panel with the same fields, all of which are required:
| Field | What to enter |
|---|---|
| Server | The SQL Server instance, for example SERVERNAME\SQLEXPRESS. |
| Database Name | The database that holds the data. |
| Database User | The login TotalQA uses to connect. This field will be ignored if Integrated Security is set to yes. |
| Password | Leave blank to keep the saved password or if Integrated Security is set to yes. |
| Integrated Security | No (standard) uses the user and password above; Yes (SSPI) uses Windows authentication. |
Here is an example:
Click Check connection in each panel to test the settings before saving.
Documents and Tasks
The right hand side of the page is for storing documents and tasks related to the specific device. None of these are required to set up the device.
- Documents: files attached to the device, such as calibration certificates. Click + Add Document to upload one.
- Upcoming Tasks: open tasks for this device, with creator, assignee, description and due date. Click + Add Task to create one.
- Completed Tasks: finished tasks, with who finished them and when.
Each list has its own search box, page size (10, 25, 50 or 100 entries) and paging controls.
Click the Sign and Update device button to save the device.
Add the Measurement Device to a Schedule
- Navigate to the "Manage Schedule Configurations" page. Select the schedule that the StarTrack3 device will be used with and click the pencil icon for Schedule Settings.
-
Expand the Dosimetry section (this may have been renamed in your settings) and then expand any of the Profile Constancy (if you have customized the section name use that instead) section. Click the Edit Test Settings button on the "Profile Constancy Photon Results" test.
-
Select the correct Equipment and Equipment Hub from the down menus.
- Select the checkbox for each energy that will be imported.
- As each energy is checked, a dropdown will be created under the "EXTERNAL ENERY" column, select the correct corresponding energy from the StarTrack3 database.
-
If not all fields will be imported, they may be deselected.
- Click the Save Machine Schedule Settings button.
- Repeat the steps above for the "Profile Constancy Electron Results" test.
Testing the Configuration
From the Schedules page, open new QA for the newly configured schedule. Click on the Get StarTrack results button. Note that the name of the button depends on the "Device Label" that was created when the measurement device was created.
A window similar to the one below will be shown:
Select the desired measurements and click the Import selected button. Compare the results to the expected results in the StarTrack3 software.
Adjust the schedule settings as needed until the imported results are as expected.